* AIRMO VOC BTX TECHNICAL DATA SHEET *
* EN14662 MCERTS CERTIFICATE *
Description
Analysis of Benzene, Toluene, Ethylbenzene, o,m,p Xylene, Styrene
A GC/FID instrument for BTEX monitoring in air, water and soil
Description :
The airmoVOC BTX is a gas chromatograph for the analysis and monitoring of trace amounts of benzene, toluene, ethylbenzene (m/p/o) xylene by FID detection.
Principal :
Automatic sampling and concentration using an absorbent trap. Desorption from the trap and injection into a metallic capillary column regulated by a temperature gradient. The H2 pressure at the column head is controlled by a piezo valve.
The detection of all compounds eluting from the column is performed by a FID detector
Characteristics :
The airmoVOC BTX is equipped with an integrated PC comprising of the Vistachrom programme. The internal calibration allows automatic validation of results. (Certified Benzene tube at ± 10%).
The instrument is robust, compact and very low maintenance and offers excellent result quality : repeatability, linearity, stability and sensibility (ppt).
MCERTS certified analyzer for benzene measurement following EN 14662-3
VISTACHROM ® software :
Chromatotec developed software system allows :
- Remote monitoring
- Full traceability through archiving of results and QC
- Set up and control of threshold alarms
- Export of data MODBUS / MGS1 / 4-20mA /0-10V
Technical Data
| Technical Specifications | |
|---|---|
| BTEX Analysis |
8 compounds with 7 peaks: Benzene, Toluene, Ethylbenzene, M+P Xylene, O Xylene, Cyclohexane* and Styrene* More VOCs in option: Naphtalene, Phenol, 1.3 Butadiene |
| Detection Limit | Benzene < 20 ppt or 0,064 µg/m3 |
| Detection Range |
0,1 to 380 µg/m3 for Benzene Standard amibient air monitoring 0 to 50 µg/m3 |
| Relative Standard Deviation |
< 0.3 % over 48 h (retention time) < 3 % over 48 h (concentration) |
| Results |
Full result storage (data and chromatogram) Embedded computer Windows® based with LCD display 128 GB of Hardware storage on SSD memory 4 USB Connecting Port Two RS-232 ports Display: 10’’ TFT Color LCD MODBUS RTU / JBUS communication protocol |
| Cycle Time |
15, 20 or 30 min (A21022 BTEX Mcerts) 30 min (A21022 BTEX-B Mcerts) 20 min (A26022) |
| Gas Supply |
H2 (FID and carrier gas): 30 ml/min (supply 2 bar ; 1/18’’ double ferrule Air (FID) : 180 ml/min (supply 3 bar ; 1/8’’double ferrule Sample supply (sample pump) 1/4’’double ferrule Pneumatic valve 90ml/commutation |