* MICRO-AIRTOXIC TECHNICAL DATA SHEET *
Description
Portable GC-PID VOC Analyzer
µ-airTOXIC is a portable GC-PID analyzer with pre-concentration for trace-level VOC monitoring. Multiple configurations cover BTEX, 1,3-butadiene and VOCs from C3 to C12, according to the selected method.
The C6-C12 versions include a PAMS configuration for 33 VOCs and a European configuration for 45 VOCs, including terpenes. Optional method libraries extend the available applications.
Applications
- Air-quality monitoring in urban areas, environmental monitoring and refinery fence lines.
- Indoor and outdoor air, industrial hygiene and CO₂ beverages.
Operating Principle and Field Deployment
An electrical six-port valve, a single adsorbent trap and a capillary column are combined with a heated PID and a 10.6 eV lamp. Temperature programming of the column oven and carrier-gas pressure/flow control support the selected analytical method.
- The portable suitcase integrates sampling, analysis, data storage, a battery and an internal nitrogen tank, with readiness for operation in less than 5 minutes.
- Integrated Wi-Fi enables control from a phone, tablet or laptop. A local touchscreen is optional.
- Vistachrom LITE provides measurement configuration, calibration, recalculation, data storage and export.
Options
- Portable permeation oven with certified permeation tubes for calibration; nitrogen generator up to 300 mL/min with an external compressor for operation and/or tank refilling, providing up to 10 hours of operation after refilling.
- Modem support and remote control; purge module for extracting VOCs from water for online BTEX analysis; additional battery.
- Rechargeable internal high-pressure carrier-gas cylinder up to 150 barg; external nitrogen bottle for up to 20 days of operation at 70 bar.
- Multiple methods with a VOC library of more than 40 compounds (µ-airTOXIC C6-C12). Biogenic measurement, including isoprene, is available, with potential co-elution of C4 or C5 ozone VOC precursors.
Application-dependent Compliance
The TSP states support for compliance with EN 14662-3:2015 for selected applications and lists EN 15267-1, EN 15267-2 and QAL1. Applicability depends on the selected configuration and application; refer to the TSP for details.
Technical Data
| Technical Specifications | |
|---|---|
| Configurations and Compounds |
BTEX: Benzene, Cyclohexane, Toluene,
Ethylbenzene, m- and p-Xylenes, Styrene and o-Xylene.
1,3-butadiene, IAA or naphthalene are optional.
C6-C12 PAMS: 33 VOCs from benzene to n-undecane. European C6-C12 configuration: 45 VOCs from benzene to n-decane, including biogenic VOCs (terpenes). Other method libraries, such as C3-C10, are optional. |
| Detection and Linearity |
Automatic-mode LDL: ≤ 10 ppt.
The TSP cites benzene or other VOCs (µ-A76022);
performance depends on compound and configuration.
Benzene ranges: 0-10 ppb, 0-100 ppb or 0-1000 ppb. Area linearity: R² > 0.995. |
| Repeatability | Relative standard deviation over 48 hours: better than 1% for retention time; better than 2% for concentration at 1 ppb. |
| Cycle Times | 10, 15, 20 or 30 minutes (µ-A73022); 30 minutes (µ-A73022-B). |
| Sampling and Carrier Gas |
Programmable sample volume:
20 to 700 mL or more.
BTEX and C6-C12 carrier-gas consumption: minimum 3 mL/min with an electric injection valve; varies with column internal diameter and application. Internal nitrogen tank: up to 15 barg. Analysis up to 22 hours at 15 bar. |
| Detector and Column |
Heated photoionization detector (PID) with a
10.6 eV lamp.
Electrical six-port valve, single adsorbent trap and capillary column. Programmed column-oven temperature gradient up to 200°C, with carrier-gas pressure/flow control. |
| Operating Temperature | 0°C to 40°C; may vary with the method used. |
| Power and Battery |
12 V DC; battery autonomy up to 8 hours,
depending on the method used.
Battery rechargeable from 230 V / 110 V, 50 Hz / 60 Hz. |
| Dimensions and Weight | Height: 410 mm; width: 340 mm; depth: 180 mm. Net weight: 7 kg. |
| Software, Storage and Communication |
Vistachrom LITE on Linux; Wi-Fi, Ethernet
and one USB port.
Standard data capacity: 16 GB. Internal data storage up to 10 years; data transfer via USB. Measurement configuration, calibration, recalculation and export; retention-time, area, mass and concentration profiles. |
| Tablet Specifications |
8.7-inch display, 1340 × 800 resolution;
Android operating system; shock protection included.
Approximately 14 hours of continuous use and approximately 335 g. Characteristics may vary. |