Chromatotec µ-airTOXIC — DiamondSci Skip to content

Chromatotec µ-airTOXIC

SKU GTIN :

* MICRO-AIRTOXIC TECHNICAL DATA SHEET *

Description

Portable GC-PID VOC Analyzer

µ-airTOXIC is a portable GC-PID analyzer with pre-concentration for trace-level VOC monitoring. Multiple configurations cover BTEX, 1,3-butadiene and VOCs from C3 to C12, according to the selected method.

The C6-C12 versions include a PAMS configuration for 33 VOCs and a European configuration for 45 VOCs, including terpenes. Optional method libraries extend the available applications.

Applications

  • Air-quality monitoring in urban areas, environmental monitoring and refinery fence lines.
  • Indoor and outdoor air, industrial hygiene and CO₂ beverages.

Operating Principle and Field Deployment

An electrical six-port valve, a single adsorbent trap and a capillary column are combined with a heated PID and a 10.6 eV lamp. Temperature programming of the column oven and carrier-gas pressure/flow control support the selected analytical method.

  • The portable suitcase integrates sampling, analysis, data storage, a battery and an internal nitrogen tank, with readiness for operation in less than 5 minutes.
  • Integrated Wi-Fi enables control from a phone, tablet or laptop. A local touchscreen is optional.
  • Vistachrom LITE provides measurement configuration, calibration, recalculation, data storage and export.

Options

  • Portable permeation oven with certified permeation tubes for calibration; nitrogen generator up to 300 mL/min with an external compressor for operation and/or tank refilling, providing up to 10 hours of operation after refilling.
  • Modem support and remote control; purge module for extracting VOCs from water for online BTEX analysis; additional battery.
  • Rechargeable internal high-pressure carrier-gas cylinder up to 150 barg; external nitrogen bottle for up to 20 days of operation at 70 bar.
  • Multiple methods with a VOC library of more than 40 compounds (µ-airTOXIC C6-C12). Biogenic measurement, including isoprene, is available, with potential co-elution of C4 or C5 ozone VOC precursors.

Application-dependent Compliance

The TSP states support for compliance with EN 14662-3:2015 for selected applications and lists EN 15267-1, EN 15267-2 and QAL1. Applicability depends on the selected configuration and application; refer to the TSP for details.

Technical Data
Technical Specifications
Configurations and Compounds BTEX: Benzene, Cyclohexane, Toluene, Ethylbenzene, m- and p-Xylenes, Styrene and o-Xylene. 1,3-butadiene, IAA or naphthalene are optional.

C6-C12 PAMS: 33 VOCs from benzene to n-undecane.

European C6-C12 configuration: 45 VOCs from benzene to n-decane, including biogenic VOCs (terpenes). Other method libraries, such as C3-C10, are optional.
Detection and Linearity Automatic-mode LDL: ≤ 10 ppt. The TSP cites benzene or other VOCs (µ-A76022); performance depends on compound and configuration.

Benzene ranges: 0-10 ppb, 0-100 ppb or 0-1000 ppb.

Area linearity: R² > 0.995.
Repeatability Relative standard deviation over 48 hours: better than 1% for retention time; better than 2% for concentration at 1 ppb.
Cycle Times 10, 15, 20 or 30 minutes (µ-A73022); 30 minutes (µ-A73022-B).
Sampling and Carrier Gas Programmable sample volume: 20 to 700 mL or more.

BTEX and C6-C12 carrier-gas consumption: minimum 3 mL/min with an electric injection valve; varies with column internal diameter and application.

Internal nitrogen tank: up to 15 barg. Analysis up to 22 hours at 15 bar.
Detector and Column Heated photoionization detector (PID) with a 10.6 eV lamp.

Electrical six-port valve, single adsorbent trap and capillary column. Programmed column-oven temperature gradient up to 200°C, with carrier-gas pressure/flow control.
Operating Temperature 0°C to 40°C; may vary with the method used.
Power and Battery 12 V DC; battery autonomy up to 8 hours, depending on the method used.

Battery rechargeable from 230 V / 110 V, 50 Hz / 60 Hz.
Dimensions and Weight Height: 410 mm; width: 340 mm; depth: 180 mm. Net weight: 7 kg.
Software, Storage and Communication Vistachrom LITE on Linux; Wi-Fi, Ethernet and one USB port.

Standard data capacity: 16 GB. Internal data storage up to 10 years; data transfer via USB.

Measurement configuration, calibration, recalculation and export; retention-time, area, mass and concentration profiles.
Tablet Specifications 8.7-inch display, 1340 × 800 resolution; Android operating system; shock protection included.

Approximately 14 hours of continuous use and approximately 335 g. Characteristics may vary.